logo
Wuhan Desheng Biochemical Technology Co., Ltd
Wuhan Desheng Biochemical Technology Co., Ltd
상품
/

상품

The True Meaning of CAPS Buffer 1135-40-6 Temperature Compensation

제품 세부 정보

원래 장소: 중국 에즈 후

브랜드 이름: DESHENG

인증: ISO9001:2008

지불 및 배송 조건

최소 주문 수량: 10g

가격: 협상 가능

포장 세부 사항: 플라스틱 병 또는 알루미늄 필름

배달 시간: 지불을받은 후 1 ~ 3 일

지불 조건: T/T, L/C, D/A, D/P, Western Union, MoneyGram, PayPal

공급 능력: 100kg/월

최상의 가격을 얻으세요
지금 연락하세요
사양
강조하다:

CAPS Buffer

,

1135-40-6

,

3- (cyclohexylamine) -1-propanesulfonic acid

모습:
화이트 크리스탈 파우더
청정:
>99.0%
MW:
221.32
공식:
C9H19NO3S
카스:
1135-40-6
이름:
모습:
화이트 크리스탈 파우더
청정:
>99.0%
MW:
221.32
공식:
C9H19NO3S
카스:
1135-40-6
이름:
설명
The True Meaning of CAPS Buffer 1135-40-6 Temperature Compensation

In the use of biological buffer CAPS, temperature compensation is a commonly mentioned but easily misunderstood concept. Many operators simply understand it as the automatic temperature compensation function on the pH meter, believing that as long as this function is turned on, the pH values measured at different temperatures can be directly compared. In fact, the true meaning of temperature compensation goes far beyond this, involving changes in the chemical equilibrium of the buffer system itself, temperature response of the measuring electrode, and temperature management strategies for experimental design. Clarifying this concept is of great significance for the correct use of CAPS buffer.


Product Name

3- (cyclohexylamine) -1-propanesulfonic acid

Chemical Abbreviation

CAPS buffer

CAS Number

1135-40-6

Product Appearance

White crystalline powder

Molecular Weight

221.32

Molecular Formula

C9H19NO3S

Density

1.19g/cm3

Melting Point

324℃

Storage Conditions

Room temperature, away from light and moisture

Standard Packaging

25KG/Cardboard bucket

Water-soluble

9 g/100 mL (20℃)

Manufacturer

Hubei New Desheng


The essence of temperature compensation in pH measurement


The response of the pH electrode is based on the Nernst equation, and its slope varies with temperature. The automatic temperature compensation function corrects the slope of the electrode response, ensuring that the potential measured by the electrode at different temperatures can be correctly converted into pH readings. But it should be clarified that this compensation does not change the actual pH value of the sample itself. When the CAPS buffer is moved from room temperature to a low-temperature environment, its true hydrogen ion concentration has already changed, and the electrode only faithfully reports this change. Therefore, automatic temperature compensation solves the problem of measurement accuracy rather than sample pH stability. Confusing the two is the most common misconception in temperature compensation cognition.


The True Meaning of CAPS Buffer 1135-40-6 Temperature Compensation 0

CAPS powder


Temperature response of CAPS buffer system itself


The acid-base dissociation constant of CAPS has a clear temperature dependence. When the temperature increases, the dissociation equilibrium shifts and the actual pH of the buffer solution changes accordingly. This change is a reflection of the inherent chemical properties of CAPS molecules and is a real change, without the concept of compensation. Understanding this is crucial as it means that any experimental design must clearly specify the operating temperature and perform pH adjustment and measurement at that temperature. Attempting to mask or eliminate this chemical change through instrument compensation is neither possible nor scientific. The correct approach is to accept and record the pH shift caused by temperature changes, and consider it in the experimental plan.


The true meaning of temperature consistency between calibration and measurement


The most direct application of temperature compensation at the operational level is to require calibration and measurement to be performed at the same temperature. If there is a temperature difference between the standard buffer used for calibration and the CAPS sample to be tested, even if the instrument is turned on for automatic compensation, it cannot eliminate the system bias introduced by the different dissociation behaviors of the standard solution and the sample. The solution is to place all liquids in the same constant temperature environment for sufficient equilibrium, and then perform calibration and measurement. For experiments that require the use of CAPS at different temperatures, it is recommended to calibrate and measure separately at the target temperature, rather than relying on the instrument to automatically convert to the target temperature after calibration at room temperature.


Temperature management strategy in workflow


Since the actual pH of CAPS varies with temperature, a reasonable temperature management strategy needs to be established during experimental design. For a constant temperature experimental platform, the problem is relatively simple, just complete calibration and adjustment at that temperature. For experiments that require variable temperature operation, such as low-temperature storage followed by temperature rise reaction, the pH value of CAPS at each key temperature point needs to be measured in advance as the basis for operation. Some experimental schemes choose to adjust the pH of CAPS at a higher temperature because the buffer capacity is more stable at this temperature. After cooling to the working temperature, it will be judged whether fine-tuning is needed based on pre experimental data. This strategy requires sufficient data support and is not a universal method.


The True Meaning of CAPS Buffer 1135-40-6 Temperature Compensation 1

Product packaging


Summary


The true meaning of CAPS buffer temperature compensation is to distinguish between measurement calibration and chemical changes, and to clarify the boundary between instrument function and operation strategy. Correct temperature management is not about eliminating changes, but about understanding the patterns of change and incorporating them into experimental design. In practical work, maintaining consistency between calibration and temperature measurement, accepting and recording the true pH shift caused by temperature in CAPS, and establishing a temperature response database for specific processes are more practical measures than relying on instrument compensation.

 

 





 

 

문의사항 보내주세요
요청사항을 보내주시면 최대한 빨리 답변하겠습니다.
보내