logo
Wuhan Desheng Biochemical Technology Co., Ltd
Wuhan Desheng Biochemical Technology Co., Ltd
produits
À la maison /

produits

Analyze HEPES 7365-45-9 compatibility interference in the detection system

Détails du produit

Lieu d'origine: HUBEI

Nom de marque: desheng

Certification: ISO9000

Numéro de modèle: CAS 7365-45-9

Conditions de paiement et d'expédition

Quantité de commande min: 20 kg

Prix: Detailed discussion

Détails d'emballage: 20 kg / boîte

Délai de livraison: 5 jours

Conditions de paiement: LC, D/P, T/T, Western Union

Capacité d'approvisionnement: 5t / semaine

Obtenez le meilleur prix
Contactez-nous maintenant
Les spécifications
Mettre en évidence:

HEPES buffer

,

4-Hydroxyethylpiperazine ethanesulfonic acid

,

7365-45-9

Apparence:
Poudre blanche
Modèle:
hépès
Produit de fabrication:
Desheng
Emballage:
500g / boîte
Apparence:
Poudre blanche
Modèle:
hépès
Produit de fabrication:
Desheng
Emballage:
500g / boîte
Définition
Analyze HEPES 7365-45-9 compatibility interference in the detection system

HEPES, as a commonly used biological buffer, although has mild chemical activity, may cause unexpected interference to the results in specific analytical tests. This kind of interference has strong concealment, but it may become a source of data bias. Fully understanding the behavior of HEPES buffer in different detection systems can help to design experimental procedures rationally and improve analytical accuracy.


Product Name

4-Hydroxyethylpiperazine ethanesulfonic acid

Chemical Abbreviation

HEPES buffer

CAS Number

7365-45-9

Product Appearance

White crystalline powder

Molecular Weight

238.305

Molecular Formula

C8H18N2O4S

Density

1.07 g/mL at 20 °C

Melting Point

234-238 °C

Storage Conditions

Room temperature, away from light and moisture

Purity

More than 99%

Standard Packaging

500g/bottle, 25kg/cardboard drum

Manufacturer

Hubei Xindesheng


Interference characteristics of protein quantification methods


In BCA and Lowry methods, the sulfonic acid and tertiary amine groups in HEPES molecules can form soluble complexes with copper ions, competitively consuming copper ions in the reaction system, resulting in a decrease in color signal. When HEPES is at the upper limit of the conventional concentration range, the measured values may produce observable lower results, and the degree of deviation is somewhat correlated with the buffer concentration. In contrast, the Bradford method is based on the binding of Coomassie Brilliant Blue to proteins and is less affected by HEPES interference, but higher concentrations of HEPES may still alter the dielectric constant of the solution, causing small fluctuations. Therefore, when performing protein quantification, it is recommended to use a standard sample with the same buffer background as the sample to draw a standard curve, or to pre desalinate the sample.


Analyze HEPES 7365-45-9 compatibility interference in the detection system 0

HEPES buffer


The influence on enzyme activity determination and coupling reaction


HEPES has a regulatory effect on the activity of some oxidoreductases and transferases. Research has shown that HEPES can participate in peroxidase catalyzed reactions, to some extent inhibiting their catalytic efficiency and affecting the determination of chemiluminescence or colorimetric endpoints based on the enzyme. In coupled enzyme assays, if hydrogen peroxide or free radical intermediates are produced in the first step of the reaction, HEPES may act as a free radical scavenger or donor, altering the reaction rate and causing the absorbance reading to deviate from the true value. In addition, HEPES has weak absorption in the short ultraviolet band. For kinetic experiments that require monitoring substrate consumption or product generation in this band, buffer blank should be deducted or a substitute with weaker ultraviolet absorption should be used.


Potential interference with fluorescence and chemiluminescence detection


Fluorescence analysis is sensitive to solution composition, and HEPES can produce background fluorescence at some excitation wavelengths, especially in the ultraviolet excitation region. Although the intensity is lower than some traditional buffering agents, it may still interfere with the detection of low abundance fluorescent probes. For time-resolved fluorescence or fluorescence polarization experiments, HEPES may alter the refractive index and viscosity of the solution, thereby affecting the polarization reading. In chemiluminescence detection, HEPES can slowly decompose to produce hydrogen peroxide in the presence of light or transition metal ions. This byproduct enhances or inhibits luminescent systems based on luminol or acridine esters, causing signal fluctuations. Using light avoidance techniques and adding an appropriate amount of metal chelating agent can help alleviate such interference.


Identification of interference and experimental countermeasures


To determine whether HEPES interferes with a specific detection system, two control strategies can be adopted. One is to set up a matrix matching control, adding a known concentration of the target substance standard to a buffer containing HEPES, and comparing whether the recovery rate falls within an acceptable deviation range. The second method is to perform dilution or displacement validation by diluting the sample with physiological saline without buffering agents and retesting. If there is a non-linear change in the results, it indicates interference from buffering agents. For experiments confirmed to be affected, it may be considered to reduce the HEPES concentration to the lowest level that meets buffering requirements, or to replace it with PIPES or EPPS with less interference within the same pH range. If HEPES must be used, it can be removed by dialysis or gel filtration before determination, but it should be noted that this operation may change the sample concentration.


Comprehensive recommendations


The compatibility issue of HEPES is not universal, but it may result in observable data offset for specific methods. Experimenters should systematically evaluate the impact of buffer type and concentration on the detection endpoint when establishing new methods, and incorporate the evaluation results into the operating procedures. The description of interference level in different literature often varies due to differences in conditions, so it is advisable to attach internal quality control samples to each batch of experiments. Through reasonable control design and necessary preprocessing, the interference effect of HEPES can be effectively controlled, thereby ensuring the authenticity and reproducibility of the analyzed data.


Analyze HEPES 7365-45-9 compatibility interference in the detection system 1

Product packaging


As a HEPES buffer manufacturer, Desheng strictly monitors product quality and sets up a quality department to inspect batches of products. Only after the packaging and quality meet the standards can they be sold from the factory. Therefore, Desheng has received unanimous praise from customers. If you have any relevant intentions, please click on the website to inquire about details and purchase!

 